Terumo Sarns 9000 Features
The Terumo Sarns 9000 is used in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures. With occluder installed in the Sarns 9000, the operator can rapidly or gradually adjust the blood flow of the venous line. The Sarns 9000 has a backup battery pack that is constantly being charged when the system is in operation. If AC power failure occurs, the battery backup will automatically supply power to the system. When adding the option to install the water mixing system, hot and cold water form a wall supply is mixed for use with cardioplegia and arterial heat exchangers. This water mixing system allows you to lower and raise the patient's temperature as needed.
- Arterial flow, oxygenation, air, and level detection have immediate automatic battery backup and function independently off the main computer.
- Arterial flow is protected by dual controls.
- Alarms, alerts, and status messages are automatically prioritized for the user.
- Allows the of use of a roller or centrifugal pump for arterial perfusion.
- Air-bubble detection.
Terumo Sarns 9000 Specifications
Console Dimensions
- Height: 25.3” (64.3 cm)
- Width: 45.3” (115.1 cm)
- Depth: 24” (61 cm)
Pump Dimensions
- Height: 9” (20.3 cm)
- Width: 8.75” (22.2 cm)
- Depth: 14” (35.6 cm)
Central Display
- Height: 10.5” (26.7 cm)
- Width: 14.1” (35.8 cm)
- Depth: 3.2” (8.1 cm)
Weight
- Roller Pump: Approx. 31 lbs (14.5 kg)
- Centrifugal Pump: 15 lbs (6.9 kg)
- Total System: 470 lbs (213 kg)
- Optional Water Mixer: Additional 60 lbs (27.2 kg)
Roller Pump
- Tubing with a maximum outer diameter of .75” (1.9 cm) and wall thickness range of 0.0625 to 0.09375” (0.16 to 0.24 cm)
- Recommended durometer range if 58-68
- Speed: At 0-200 RPM, display accuracy is +/- 2 RPM; Resolution is 1 RPM
- From 200 RPM to a value within 248-258 RPM, display accuracy is +/- 1% of actual speed; Resolution is 1 RPM
- Flow Rate: At 0.00-9.99 L/min. Displays 999 when > 9.99 L/min
- Operating Modes: Continuous or Asynchronous Pulsatile
Centrifugal Pump
- Speed display range is 0-9999 RPM
- Actual speed range is 0-3600 RPM
- Flow sensing range is 0.7 to 0.7 L/min
- Flow sensing accuract is +/- 0.1 L/min within a range of 0.7 - 1.0 L/min and +/- 10% accuracy within a range of 1.0 - 7.0 L/min
- Outflow capacity is 0-9.9 L/min
- Outflow pressure is 0-700 mmHG
- Flow display range with a sensor for 0.375” (0.95 cm) ID tubing is 0.50 - 9.90 L/min in 0.1 increments. Displays 999 when > 9.90 L/min
- Operating Modes: Continuous flow or Asynchronous Pulsatile